• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM; SUTURE MEDIATED CLOSURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM; SUTURE MEDIATED CLOSURE Back to Search Results
Catalog Number 12773-03
Device Problems Improper or Incorrect Procedure or Method (2017); Obstruction of Flow (2423); Malposition of Device (2616)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/06/2024
Event Type  Injury  
Event Description
It was reported that this was a closure of a very deep right common femoral vein with a prostyle device using the pre-close technique via a 6fr sheath hole prior to a left atrial appendage occlusion procedure.Reportedly, there was no bleed back from the marker lumen.It was viewed under fluoroscopy, and it was judged that the device was in the vein enough, therefore the suture was deployed and pre-placed.The sheath was upsized to 16f, and the procedure was completed.After advancing the knot to the vessel wall, bleeding continued from the access site.It was noted that the suture was deployed in soft tissue and not the vein.A figure of eight suture was applied to achieve hemostasis.There was no reported adverse patient sequela and no reported clinically significant delay in the procedure or therapy.No additional information was provided.
 
Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.H6: medical device problem code 2017 clarifier- failure to follow steps/ instructions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM
Type of Device
SUTURE MEDIATED CLOSURE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18984035
MDR Text Key338658176
Report Number2024168-2024-03731
Device Sequence Number1
Product Code MGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number12773-03
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/06/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age57 YR
Patient SexFemale
Patient Weight129 KG
-
-