ABBOTT DIABETES CARE INC FREESTYLE LIBRELINK; DATA MANAGEMENT SYSTEM Back to Search Results
Model Number 71732-01
Device Problems Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591); Application Program Problem (2880)
Patient Problems Vomiting (2144); Loss of consciousness (2418)
Event Date 02/24/2024
Event Type  Injury  
Manufacturer Narrative
The customer reported a "replace sensor" error message with the freestyle libre application.Abbott diabetes care (adc) attempted to replicate the reported issue and the reported configuration of device huawei p smart is not compatible with the freestyle librelink application.The compatibility guide is available to the customer on the abbott diabetes care website.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time, product has not yet been returned.An extended investigation has been performed for the reported complaint and determined that there was no indication that the product did not meet specification.The dhrs (device history record) for the libre sensor and sensor kit were reviewed, and the dhrs showed the libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
An error message was reported with the adc device in use with huawei p smart 2019 phone with android operating system version 10.A customer reported a "replace sensor" error message with the adc device and was unable to obtain readings.As a result, the customer experienced vomiting and a loss of consciousness and was administered unspecified intravenous treatment by a healthcare professional (hcp).There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRELINK
Type of Device
DATA MANAGEMENT SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda 94502-7001
5107495297
MDR Report Key18984574
MDR Text Key338658439
Report Number2954323-2024-10074
Device Sequence Number1
Product Code PZE
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
P160030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71732-01
Was Device Available for Evaluation? No
Date Manufacturer Received03/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2024
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? Yes
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-