The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd drive pcb.In addition, our technician confirmed that the lg connector fluid damage, the insertion flexible tube perforated, the remote control buttons perforated, the insertion flexible tube buckled, the lg connector corroded, the insertion flexible tube leak, the remote control buttons leak, the root brace rubber (insertion flexible tube) cut, the distal body worn out, the lcb (light carrying bundle) defective, the operation channel (primary) resistance, and the suction channel kink; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|