Model Number CF-EZ1500DL
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/10/2024
Event Type  malfunction  
Manufacturer Narrative
The issue was not confirmed, as the scope was not microbiologically tested by olympus.The evaluation of the event is ongoing and follow up with the customer is currently being performed.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the colonovideoscope tested positive twice for an unexpected contamination.The issue was found during routine cultures of the scope.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacture's final investigation.Despite good faith attempts the user culture results and cleaning disinfection and sterilization (cds) processes were not shared.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation and because the subject device was not returned, the reported event could not be confirmed and a root cause could not be determined.The following is included in the device ifu: "an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COLONOVIDEOSCOPE
Type of Device
COLONOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18985759
MDR Text Key338693221
Report Number9610595-2024-06440
Device Sequence Number1
Product Code FDF
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCF-EZ1500DL
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/10/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-