• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC CORPORATION HEARTMATE II LVAS IMPLANT KIT (WITH SEALED GRAFTS); VENTRICULAR (ASSIST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC CORPORATION HEARTMATE II LVAS IMPLANT KIT (WITH SEALED GRAFTS); VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number 106015
Device Problems Mechanical Problem (1384); Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/07/2024
Event Type  malfunction  
Manufacturer Narrative
Historically related mfr 2916596-2019-03994 covering the full-length driveline replacement in the past; 2916596-2020-03532 covering short to shield.No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
 
Event Description
It was reported that the patient was seen in clinic with pump stop alarms, low speed advisories, and low flow alarms.Event log files were submitted for review.The event log confirmed that pump speed dropped below the low-speed limit and captured pump stops on 07mar2024 and 08mar2024.These events occurred on battery power and appeared to be caused by a phase to phase short.It was noted that the patient had a full-length driveline replacement previously, so another driveline repair was not possible.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE II LVAS IMPLANT KIT (WITH SEALED GRAFTS)
Type of Device
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
lindsey sallese
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key18986067
MDR Text Key338699296
Report Number2916596-2024-01800
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00813024011224
UDI-Public00813024011224
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P060040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number106015
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/19/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age46 YR
Patient SexMale
Patient Weight108 KG
-
-