ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA MP ICD; No Match Back to Search Results
Model Number CD3369-40Q
Device Problems Failure to Capture (1081); High impedance (1291); Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2023
Event Type  Injury  
Event Description
It was reported that high out of range pacing impedance and high out of range high voltage (hv) impedance was observed on a remote transmission.The patient was brought into the clinic to be evaluated further.In the clinic, it was confirmed all three leads had an out of range impedance, no capture, and no sensing.Therapies were disabled and pacing was turned off in the clinic.X-rays showed the leads were in the original implant location.The patient was brought into the lab to test leads, and all leads were tested within normal limits with no abnormalities present.A new generator was implanted and the chronic leads connected to the new generator in working condition.A device/header malfunction was suspected.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
No Match
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18986298
MDR Text Key338703838
Report Number2017865-2024-37020
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508377
UDI-Public05414734508377
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Model NumberCD3369-40Q
Device Lot NumberA000075269
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/23/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OPTISURE; QUARTET; TENDRIL
Patient Outcome(s) Required Intervention;
Patient Age72 YR
Patient SexMale
-
-