Model Number 48-40-00Z |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2024 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland received a report that a s5 pump gave a 'motor monitoring /control failed' error message during procedure.Pump was working and the customer could clear the error message, but it returned every 15 minutes.There was no patient injury.
|
|
Manufacturer Narrative
|
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 pump.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Manufacturer Narrative
|
A livanova field service engineer was dispatched to the customer site and was able to duplicate issue.To solve the issue, pump head and hkr board were replaced and all system flashed to be compatible.The device was tested and returned to customer.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
See initial report.
|
|