• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORAX MEDICAL, INC. 1.5T LINX, 14B; ANTI-REFLUX IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORAX MEDICAL, INC. 1.5T LINX, 14B; ANTI-REFLUX IMPLANT Back to Search Results
Catalog Number LXMC14
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Cramp(s) /Muscle Spasm(s) (4521)
Event Date 03/13/2024
Event Type  Injury  
Manufacturer Narrative
(b)(4) date sent: 3/27/2024 attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.Additional information was requested, and the following was obtained: on what date did the implant take place? on (b)(6) 2023.On what date did the explant take place? on (b)(6) 2024.What is the device lot #? 30089 does the patient have any of the allergies to metals? if so, what test have been done to test for metal allergies.None is the patient currently taking currently taking steroids / immunization drugs? none did the patient have any pre-existing dysphagia or other conditions (other than gerd)? hypertension, schatzki's ring was there any hiatal or crural repair done at the same time as the implant? yes was mesh used at time of implant? yes what was the reason for removal of the linx device? persistent esophageal spasm at the time of removal, was the device found in the correct position/geometry at the time of removal? yes have the symptoms resolved since the device was explanted? unknown at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that a linx device was explanted for unknown reasons.Patient current status was not provided to the sales rep.
 
Manufacturer Narrative
(b)(4).Date sent: 4/11/2024.Investigation summary: overall review of the device function and dimensions show no anomalies from a device that has been reasonably changed as part of the explant procedure.Visual analysis was consistent with an explanted device, tooling marks were noted in some beads.Link length and tensile force were found to meet the applicable specifications.Overall, no analysis conclusions relevant to the patient experience were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5T LINX, 14B
Type of Device
ANTI-REFLUX IMPLANT
Manufacturer (Section D)
TORAX MEDICAL, INC.
4545 creek road
cincinnati OH
Manufacturer Contact
kate karberg
4545 creek rd.
cincinnati, OH 45242
3035526892
MDR Report Key18987220
MDR Text Key338719294
Report Number3008766073-2024-00051
Device Sequence Number1
Product Code LEI
UDI-Device Identifier00855106005349
UDI-Public00855106005349
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberLXMC14
Device Lot Number30089
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received04/04/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-