Catalog Number 530.710 |
Device Problems
Mechanical Problem (1384); Physical Resistance/Sticking (4012); Difficult or Delayed Separation (4044)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2024 |
Event Type
malfunction
|
Event Description
|
It was reported from japan that during service and evaluation, it was determined that the battery oscillator device had component damage, moving parts of the trigger of the device did not move smoothly, the device could not secure/lock the cutter, and the device would not run.It was noted that there were dents and scratches on blade gripping area, dial knob spinning, poor trigger return, saw blade cannot be installed.It was further determined that the device failed pretest for general condition, check the quick coupling for saw blades, check for sticky trigger, check the function of the device, and check oscillation frequency with frequency meter.It was noted in the service order that the locking knob did not function, and it was difficult to remove the blade.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2024.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that moving parts of the trigger of the device did not move smoothly, and the device could not secure/lock the cutter.Therefore, the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to wear.Udi (b)(4).
|
|
Search Alerts/Recalls
|
|
|