Sensor (b)(6) has been returned and investigated.The sensor plug is properly seated and no physical damage was observed.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 5 (indicating normal termination).Visual inspection has been performed on the sensor plug assembly, no failure modes were observed.Perform sim vivo testing (simulation of the electrical signal produced by the sensor tail) and poise voltage testing.All results were within specification.Therefore, issue is not confirmed.No malfunction or product deficiency was identified.Dhrs (device history review) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and freestyle libre sensor kit passed all tests prior to release.All pertinent information available to abbott diabetes care has been submitted.
|