• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM Back to Search Results
Catalog Number AB2000
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/14/2024
Event Type  malfunction  
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation (procept) became aware that during jet alignment, the aquabeam robotic system generated a "e21 - motorpack error".Multiple troubleshooting steps to clear the error were unsuccessful.A second aquabeam handpiece was used which resolved the issue.The reported event caused a surgical delay of over 20 minutes.There were no adverse health consequences to the patient due to this event.
 
Manufacturer Narrative
Root cause of the reported event has not yet been established.Investigation by manufacturer is currently in process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Manufacturer Narrative
The aquabeam handpiece was not returned for investigation of the reported event.Three (3) good faith efforts were made to retrieve the device without success.The treatment log files were reviewed.No occurences of "e21 - motorpack" errors were observed; however, multiple "e22 - motorpack" and "e23 - motorpack" errors were observed to have occurred.The root cause of the reported event was unable to be established as the device was not returned for investigation.A review of the device history record (dhr) for aquabeam robotic system/serial number (b)(6) and the aquabeam handpiece / lot number 23c10522 was performed, which confirmed that there was one (1) non-conformance issued to this lot during the manufacturing process that could potentially be related to the reported event.The affected units within the lot were segregated and reworked to address the non-conformances.The handpiece passed final inspection prior to release for distribution.The current user manual sj-um0101-00 rev.B, um, aquabeam robotic system user manual, us, baytech was reviewed and states: table 5: system detected errors and faults: e21 - motorpack error.Release foot pedal and click x.If error persists, replace motorpack.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
150 baytech drive
san jose CA 95134
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
150 baytech drive
san jose CA 95134
Manufacturer Contact
doria esquivel
150 baytech drive
san jose, CA 95134
MDR Report Key18987941
MDR Text Key338762959
Report Number3012977056-2024-00093
Device Sequence Number1
Product Code PZP
UDI-Device IdentifierB614AB20001
UDI-Public+B614AB20001/$$722C05071/16D20221222B
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAB2000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/24/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/22/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-