• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC CORPORATION HEARTMATE 3 LVAS IMPLANT KIT; VENTRICULAR (ASSIST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC CORPORATION HEARTMATE 3 LVAS IMPLANT KIT; VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number 106524US
Device Problems Mechanical Problem (1384); Infusion or Flow Problem (2964)
Patient Problem Thrombosis/Thrombus (4440)
Event Date 03/04/2024
Event Type  Injury  
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
 
Event Description
It was reported that the patient had an outflow track thrombus.Log files were submitted for review and captured low flow events on (b)(6).The calculated flow appears to have fluctuated below the alarm threshold of 2.5 lpm.Some of the low flow events were not sustained for long enough (at least 10 seconds) to activate the audible alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE 3 LVAS IMPLANT KIT
Type of Device
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
lindsey sallese
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key18988379
MDR Text Key338730894
Report Number2916596-2024-01569
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00813024013297
UDI-Public00813024013297
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/10/2021
Device Model Number106524US
Device Lot Number6781842
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/07/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/17/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age56 YR
Patient SexMale
Patient Weight76 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-