This supplemental report is being submitted to provide the device inspection results, the results of the legal manufacturer's final investigation, the associated h6 coding, and update to field h3.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 2 years since the subject device was manufactured.An olympus field service engineer (fse) went on site.The device was inspected it was found that when running on program 1, the cycle took about 40 minutes to complete.The external water filter output pressure gauge readings were noted as 0.45 marking low at 20 psi during rinse.The fse replaced both external and internal filters, the pressure increased to around 50 psi and the cycle took about 30 minutes to complete.This is what the customer has been doing to improve the cycle time.The fse noticed the filter was pink in color indicating the issue was coming from the quality of the water supply delivered to all the machines.The fse discussed the water quality and requirements.The customer was using a water softener for the water supply.The customer will contact the city to inquire about the water quality and to confirm the correct water softener was being used.The device passed functional testing.Based on the results of the investigation and the information provided, root cause was likely be due to the clogged/discolored water filter, which decreased the volume of water supplied to the reprocessing bin and therefore, the total time of the reprocessing cycle was increased.The issue can be prevented by following the instructions provided in: chapter 8 replacement of consumable items, section 8.4 replacing the water filter (maj-824 or maj-2318).Olympus will continue to monitor field performance for this device.
|