• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HARH36
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/25/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: 3/27/2024.D4 batch #: a9ce4a.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the device was returned with the tissue pad detached and not returned.The device was connected to a test hand piece and a gen11, and the device did activate during functional testing.The device was disassembled to verify the internal components and no anomalies were found.   based on the condition of the tissue pad, a probable cause of this damage is that the clamp of the device may have been closed and the instrument activated without tissue present.Care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad.The resulting damage contributes to the removal of the pad from the clamp arm.The cleaning of the pad, not in accordance with the ifu, can also result in removal of the pad during use.As part of the quality process all devices are manufactured, inspected, and released to approved specifications.  a manufacturing record evaluation was performed for the finished device lot/batch a9ce4a and no non-conformances were identified.
 
Event Description
It was reported that during a radical prostatectomy the tissue pad melted.Changed to another device to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key18988994
MDR Text Key338736353
Report Number3005075853-2024-02451
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10705036014621
UDI-Public10705036014621
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K132612
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHARH36
Device Lot NumberA9CE4A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2024
Date Manufacturer Received03/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/04/2023
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
HANDPIECE AND GENERATOR.
-
-