A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 8 years since the subject device was manufactured.Based on the results of the investigation, it is likely that plastic distal end cover (c-cover) cracked due to physical stress from impact.However, a definitive root cause could not be determined.The event can be detected by following the instructions for use for inspection pf the endoscope: inspection of the forceps elevator mechanism: inspection for smooth operation: visually confirm that the portion of the elevator wire extending from the distal end of the endoscope is not broken, frayed, or bent (see figure 3.5).If any damage (broken, frayed, or bent portion) is observed on the elevator wire as shown in figure 3.5, do not use the endoscope.Olympus will continue to monitor field performance for this device.
|