Model Number TJF-150
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/12/2024
Event Type  malfunction  
Event Description
It was reported the duodeovideoscope was cracked on the distal end.The issue occurred during preparation for use.The intended therapeutic stone removal procedure was completed using a similar device.There were no reports of patient harm.
 
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 8 years since the subject device was manufactured.Based on the results of the investigation, it is likely that plastic distal end cover (c-cover) cracked due to physical stress from impact.However, a definitive root cause could not be determined.The event can be detected by following the instructions for use for inspection pf the endoscope: inspection of the forceps elevator mechanism: inspection for smooth operation: visually confirm that the portion of the elevator wire extending from the distal end of the endoscope is not broken, frayed, or bent (see figure 3.5).If any damage (broken, frayed, or bent portion) is observed on the elevator wire as shown in figure 3.5, do not use the endoscope.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18989704
MDR Text Key338757878
Report Number9610595-2024-06495
Device Sequence Number1
Product Code FDT
UDI-Device Identifier04953170307553
UDI-Public04953170307553
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K954451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTJF-150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/09/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/02/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-