ABBOTT QUADRA ASSURA MP ICD; No Match Back to Search Results
Model Number CD3371-40QC
Device Problems Incorrect Interpretation of Signal (1543); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/12/2024
Event Type  Injury  
Event Description
During a clinic follow-up, inappropriate anti-tachycardia pacing (atp) was delivered by the device due to an incorrect interpretation of signal.The device was reprogrammed to resolve the event.The patient was stable and will continue to be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
No Match
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18989793
MDR Text Key338742355
Report Number2017865-2024-37057
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2023
Device Model NumberCD3371-40QC
Device Lot NumberP000116416
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/12/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/29/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-