Model Number L311 |
Device Problem
Under-Sensing (1661)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/08/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the patient with this pacemaker device exhibited undersensing.Upon review, there were irregular amplitude, however the rhythm was consistent with the ventricular fibrillation (vf).The rhythm was undersensed at times so the ventricular pacing was delivered.At the end of the event, the vf persisted, and no other ventricular events were recorded after that.The presenting electrogram (egm) showed no deflection which indicated capture.The health care professional (hcp) called the patient, but no response received.This device remains in service.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Correction to b5 describe event or problem.Added additional details.
|
|
Event Description
|
It was reported that the patient with this pacemaker device exhibited undersensing.Upon review, there were irregular amplitude, however the rhythm was consistent with the ventricular fibrillation (vf).The rhythm was undersensed at times so the ventricular pacing was delivered.At the end of the event, the vf persisted, and no other ventricular events were recorded after that.The presenting electrogram (egm) showed no deflection which indicated capture.The health care professional (hcp) called the patient, but no response received.This device remains in service.No adverse patient effects were reported.This supplemental report is being filed as the undersensing allegation appears to be due to programmed sensitivity, and to clarify the pacemaker would not treat ventricular fibrillation (vf).This device operated as expected.
|
|
Search Alerts/Recalls
|
|