MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EMBLEM S-ICD; IMPLANTABLE ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EMBLEM S-ICD; IMPLANTABLE ELECTRODE Back to Search Results
Model Number 3501
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Shock from Patient Lead(s) (3162)
Event Date 10/07/2023
Event Type  Injury  
Event Description
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) exhibited noise that was stored as an atrial fibrillation (af) event.Additionally, there was an untreated event.Both times the device was programmed in alternate vector 1x.A chest x-ray was performed, and everything was normal.All three vectors were viewed and there was no evidence of any issues.Auto setup failed in primary due to a few good beats, secondary passed, and alternate passed and was chosen.Technical services (ts) discussed possible causes.The field representative will discuss with the health care professional (hcp).Additional information was received which reported this patient received multiple inappropriate shocks due to noise that was oversensed.Subsequently, the system was explanted and replaced.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM S-ICD
Type of Device
IMPLANTABLE ELECTRODE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18989805
MDR Text Key338742528
Report Number2124215-2024-18616
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526599200
UDI-Public00802526599200
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110042/S077
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/19/2020
Device Model Number3501
Device Catalogue Number3501
Device Lot Number151614
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/25/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/19/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age53 YR
Patient SexMale
-
-