• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSCOR INC. SAFESHEATH® II

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSCOR INC. SAFESHEATH® II Back to Search Results
Model Number SS6
Device Problem Material Too Rigid or Stiff (1544)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/18/2023
Event Type  malfunction  
Event Description
When attempting to peel away the introducer it seemed more 'rigid' than normal.Had to use scissors to cut, it happened with two introducers from same box (model and lot number).The issue was resolved and completed successfully by using another accessory (introducer).No indication of product returning for evaluation.No adverse patient effects were reported.
 
Manufacturer Narrative
No product was provided for analysis.When attempting to peel away the introducer it seemed more 'rigid' than normal.Had to use scissors to cut, it happened with two introducers from same box (model and lot number).The issue was resolved and completed successfully by using another accessory (introducer).There was no adverse event with the patient.The device history records were reviewed to confirm that the device passed all applicable in-process and final inspections.No product was returned to oscor inc.For evaluation; however, a complaint notification was provided.Without the return of the actual device in question for evaluation, oscor inc.Is unable to determine the exact cause for this incident.At this time, it is not possible to assign a definitive root cause for the event as reported.The complaint is non-verifiable as the product was not returned for evaluation.The instructions for use (ifu) adelante safesheath ii, informs the user: sharply snap the tabs of valve housing in a plane perpendicular to the long axis of the sheath to split the valve and peel sheath apart while withdrawing from the vessel.No further follow-up is required.The event will be re-evaluated if additional information becomes available.Oscor will continue to monitor this event type and risk.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFESHEATH® II
Type of Device
SAFESHEATH® II
Manufacturer (Section D)
OSCOR INC.
3816 desoto blvd.
palm harbor FL 34683
Manufacturer (Section G)
OSCOR INC.
3816 desoto blvd.
palm harbor FL 34683
Manufacturer Contact
daniel naut
3816 desoto blvd.
palm harbor, FL 34683
7279372511
MDR Report Key18990209
MDR Text Key338773731
Report Number1035166-2024-00019
Device Sequence Number1
Product Code DYB
UDI-Device Identifier30891492002211
UDI-Public30891492002211
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSS6
Device Catalogue NumberSS6
Device Lot NumberDP-19676
Was Device Available for Evaluation? No
Date Manufacturer Received03/05/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/07/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-