Model Number 304-20 |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/04/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
|
|
Event Description
|
At a prophylactic battery replacement, diagnostics in pre-op were good.During the replacement procedure, it appeared that the lead body was highly coiled, almost in a knot.The new generator was connected and a diagnostic was run.High impedance was seen, so the lead pin was removed, cleaned, and re-connected.High impedance was seen again.The lead pin was unseated a second time and re-inserted.High impedance was seen a third time.The old generator was then connected and high impedance was seen.The lead body was examined and surgeon noted coiling and knotting of the lead body.There could possibly be a microfracture that was not visible.Discussion with neurologist and family led to decision to not replace lead.The new generator was implanted and turned on with final impedance reading still high.The explanted generator is not available for return to the manufacturer.It was later reported that per the patient's surgeon that there was likely some manipulation going on by patient per review of x-rays.No other relevant information has been received to date.No additional, relevant surgical intervention has occurred to date.
|
|
Search Alerts/Recalls
|
|