Per visual inspection: no physical damage to cartridge holder or inpen front and back shell was noted.The leadscrew was received 1/4 it's travel.Several attempts were made to pair inpen, every time app displayed ¿inpen not found¿.The inpen does not pair with commercial mobile app.Inpen did transmit to manufacturing app.However, unable to perform baseline/wireless functionality and displacement dose accuracy due to not pairing to commercial mobile app.And difficult to dial/dose.Performed encoder bond investigation and found that the encoder pattern wheel tabs rotating and traveling off the keyed slots of dose nut guides.Encoder pattern wheel should never rotate.This causes an unexpected travel of the encoder pattern wheel creating resistance to dial and dispensing.Abrasions only at the top section of the dose nut.It was determined from destructive analysis that the difficult to dial/dose was caused by a pattern wheel misalignment due to an encoder base bond failure.This can affect insulin delivery.Dose log/report data inaccuracy confirmed and difficult to dial/dose confirmed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|