• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COMPANION MEDICAL INC INPEN MMT-105ELBLNA ELI LILY BLUE; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COMPANION MEDICAL INC INPEN MMT-105ELBLNA ELI LILY BLUE; SYRINGE, PISTON Back to Search Results
Model Number MMT-105ELBLNA
Device Problems Computer Software Problem (1112); Material Too Rigid or Stiff (1544)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/01/2024
Event Type  malfunction  
Manufacturer Narrative
Per visual inspection: no physical damage to cartridge holder or inpen front and back shell was noted.The leadscrew was received 1/4 it's travel.Several attempts were made to pair inpen, every time app displayed ¿inpen not found¿.The inpen does not pair with commercial mobile app.Inpen did transmit to manufacturing app.However, unable to perform baseline/wireless functionality and displacement dose accuracy due to not pairing to commercial mobile app.And difficult to dial/dose.Performed encoder bond investigation and found that the encoder pattern wheel tabs rotating and traveling off the keyed slots of dose nut guides.Encoder pattern wheel should never rotate.This causes an unexpected travel of the encoder pattern wheel creating resistance to dial and dispensing.Abrasions only at the top section of the dose nut.It was determined from destructive analysis that the difficult to dial/dose was caused by a pattern wheel misalignment due to an encoder base bond failure.This can affect insulin delivery.Dose log/report data inaccuracy confirmed and difficult to dial/dose confirmed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported to medtronic minimed that the customer experienced dose log/report data inaccuracy, difficult to dial/dose.The customer reported no adverse event.The event involved product(s) mmt-105elblna.Customer reported dose knob/dial difficult to turn.Inpen replacement initiated.Customer reported dose knob/dial difficult to turn.No harm requiring medical intervention was reported.Mmt-105elblna was requested and device was returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105ELBLNA ELI LILY BLUE
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer (Section G)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer Contact
lusine boyadzhyan
1223 world trade drive, 100
san diego, CA 92128
MDR Report Key18992350
MDR Text Key339279315
Report Number3012822846-2024-00387
Device Sequence Number1
Product Code FMF
UDI-Device Identifier0000108620880003
UDI-Public(01)0000108620880003(17)231109(10)B0838
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201337
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/09/2023
Device Model NumberMMT-105ELBLNA
Device Catalogue NumberMMT-105ELBLNA
Device Lot NumberB0838
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2024
Date Manufacturer Received03/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Age82 YR
Patient SexFemale
Patient Weight112 KG
-
-