• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CADD SOLIS VIP PUMP; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CADD SOLIS VIP PUMP; PUMP, INFUSION Back to Search Results
Model Number 2120
Device Problem Insufficient Flow or Under Infusion (2182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2024
Event Type  malfunction  
Event Description
It was reported that pump stopped three hours earlier than it was supposed to.11mls left in the bag, but pump read 0mls.There was usually overfill of about 9.2mls with the patient's blincyto pumps.Pump was programmed correctly but would not continue to infuse.The pumped alarmed and message says complete.There was a closed system transfer device used.The pump was used to prime the extension tubing.The patient was able to a new pump put on with no issues.Drug infused: blincyto.Volume to be infused: 100.8mls.Volume remaining: 11mls.Reservoir: 0mls, infusion rate: 0.6ml/hr.Air detector and upstream sensor were on.Length of infusion was 7 days pump.No adverse patient effects were reported by the customer.
 
Manufacturer Narrative
H3: device not received by manufacturer.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD SOLIS VIP PUMP
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan ln n
minneapolis, MN 55442
2247062300
MDR Report Key18992885
MDR Text Key338814209
Report Number3012307300-2024-01787
Device Sequence Number1
Product Code FRN
UDI-Device Identifier15019517150001
UDI-Public15019517150001
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111275
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number2120
Device Catalogue Number21-2120-0105-01
Was Device Available for Evaluation? No
Date Manufacturer Received02/29/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/22/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
BLINCYTO; CST; EXTENSION TUBING
-
-