US SURGICAL PUERTO RICO ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number EGIA60AMT
Device Problems Misfire (2532); Failure to Form Staple (2579); Detachment of Device or Device Component (2907)
Patient Problem Tissue Breakdown (2681)
Event Date 02/29/2024
Event Type  malfunction  
Event Description
According to the reporter, in a laparoscopic bariatric sleeve gastroplasty, during gastric stapling, the device fired but the doctor was unable to perform full stapling.The reload partially fired.Only the beginning part were stapled, some of the staples stayed open in the cavity while the rest stayed in the reload and did not cut the tissue.Some of the fired staples were also malformed.Four more reloads from another lot was used but the same issue occurred.The stapler was switched to a powered stapler and the stapling was completed.The malformed staples were removed and new stapling was carried out with another reload.The staples that fell into the cavity were removed by using forceps and sterile gauze.The surgical time was extended for 30 minutes or more due to the replacement of the faulty devices.
 
Manufacturer Narrative
D10 concomitant product: egia45amt egia 45 artic med thick sulu (lot#: p2m0197); egia60amt egia 60 artic med thick sulu (lot#: p2l0397); egia60amt egia 60 artic med thick sulu (lot#: p2l0397); egia60amt egia 60 artic med thick sulu (lot#: p2l0397); egiauxl endogia ultra univ xl stapler (lot#: p2l0508) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Manufacturer Narrative
Additional information: h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.A video was also provided.Visual inspection noted malformed staples being retrieved from a patient during a surgical procedure.The first video showed grasping tools being used to show the staple line which has malformed staples.The second video showed the reload clamped on an organ and the grasping tool moving around during a surgical procedure.It was reported that the reload partially fired.The reported issue could not be confirmed.The most likely cause could not be established from the information available.It was also reported that a component disengaged and some staples were also malformed.The reported issues were confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18993245
MDR Text Key338914426
Report Number2647580-2024-01410
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10884523003215
UDI-Public10884523003215
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K111825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEGIA60AMT
Device Catalogue NumberEGIA60AMT
Device Lot NumberP2L0397
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/17/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SEE NOTE ON H11.
Patient Age42 YR
Patient SexMale
-
-