Device evaluated by manufacturer: a visual, tactile and microscopic examination was performed on the returned device.An examination of the hypotube identified that break had occurred in the hypotube and the break was located at 59cm proximal from the tip section of the device.The proximal section of the hypotube break, including the hub manifold, was not returned for analysis.A microscopic examination the hypotube break site identified that the break is consistent with excessive forces having been applied to the hypotube.A visual, tactile and microscopic examination of the distal shaft polymer extrusion identified no damages.An examination of the balloon identified no damages.The balloon did not appear to have been subjected to any positive pressure and the balloon was tightly folded.An examination of the crimped stent identified no damages.The stent was crimped between the proximal and distal markerbands.Inspection of the remainder of the device presented no other damage or irregularities.
|