As reported by the field clinical specialist (fcs), during a transfemoral tavr procedure with a 29mm sapien 3 valve, the balloon on the 29mm commander delivery system ruptured during valve deployment at full inflation.There were no issues with the valve.However, when trying to remove the delivery system, the operator was unable to get the delivery system into the 16 fr esheath+.It was decided to remove the devices as a unit, but they were unable to get it through the tight iliac.A cut down of the femoral was performed and they were able to remove the esheath+ and delivery system.The artery was surgically closed.Per report, there was mild/moderate leaflet and sinotubular junction calcium.
|