It was reported that right ventricular (rv) defibrillation lead exhibited a high out-of-range shock impedance measurement of 140 ohms.The shock impedance measurement at implant was 83 ohms, and measurements over the last year were in the 110s-120s ohms range.Evidence suggests lead calcification likely the cause of the gradually increasing shock impedance measurements over the last year.Technical services (ts) discussed troubleshooting options and programming considerations, such as performing commanded shocks.This rv lead remains in service.No adverse patient effects and no interventions were reported at this time.
|