Model Number CLV-190
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/14/2024
Event Type  malfunction  
Event Description
It was reported, the xenon light source light control from the processor was not working and the brightness adjustment would not adjust brightness to make tissue distinguishable.It is unknown when the issue occurred.The procedure was completed using the same device.There were no reports of patient harm.
 
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Additional information was added to fields d10, h3, h4 and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 2 years since the subject device was manufactured.The olympus technical assistance center (tac) performed troubleshooting with the customer, and it was determined that the maj-1941 light control cable was loose.Reconnecting the cable resolved the issue.Three attempts were performed to obtain additional information regarding the patient and event information, but no response to the questions were provided.The customer did confirm that the issue was resolved with the tac troubleshooting provided.Based on the results of the investigation, it is presumed that the reported issues resulted from the light control cable maj-1941 being loose.However, a specific root cause could not be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18995731
MDR Text Key339015585
Report Number3002808148-2024-02944
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/14/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
MAJ-1941-LIGHT SOURCE CABLE
-
-