MEDTRONIC MED REL MEDTRONIC PUERTO RICO ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number 37612 |
Device Problems
Connection Problem (2900); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/22/2024 |
Event Type
malfunction
|
Event Description
|
Patient representative said recharger wasn't working right and not charging correctly for about a month.The rep had a bent desktop charger connector pin.An email was sent to repair to replace the desktop charger.No symptoms were reported.
|
|
Manufacturer Narrative
|
Section d references the main component of the system.Other medical products in use during the event include: brand name restore; product id 37761 (serial: (b)(6)); product type: 0213-recharger.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Manufacturer Narrative
|
H3.Analysis of the desktop charger (s/n (b)(6)) found the cord was frayed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|