Model Number 402265
Device Problems High impedance (1291); Ambient Noise Problem (2877)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/22/2024
Event Type  malfunction  
Event Description
Shock impedance shows over 150 ohms on this lead.This is a sudden change from the previous values around 45 ohms.Noise is also evident on the far field channel in multiple recordings.Full lead evaluation was recommended.Lead remains implanted.No adverse patient events were reported.Should additional information become available, this file will be updated.
 
Manufacturer Narrative
The lead is currently not available for analysis.No conclusion can be drawn based on available information at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Manufacturer Narrative
(b)(6) 2024 - lead explanted.The lead is currently not available for analysis.No conclusion can be drawn based on available information at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLEXA PROMRI S 60
Type of Device
ICD LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
DE  12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key18996382
MDR Text Key338989721
Report Number1028232-2024-01771
Device Sequence Number1
Product Code NVY
UDI-Device Identifier04035479140014
UDI-Public(01)04035479140014(17)250930
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number402265
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/22/2024
Date Device Manufactured09/15/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age74 YR
Patient SexMale
-
-