Model Number 2120 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
B3: unknown.One device was received for evaluation.Visual inspection found the tamper seal to be broken.Functional testing was able to verify and duplicate the reported issue.It was determined that the out of specification battery compartment was the root cause and replaced.The service history review identified no indication that the complaint was related to a service of the device within the review period.
|
|
Event Description
|
It was reported that the device's battery compartment was burnt.The product fault occurred during testing, there was no patient involvement.
|
|