• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CADD SOLIS VIP PUMP; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CADD SOLIS VIP PUMP; PUMP, INFUSION Back to Search Results
Model Number 2120
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/01/2024
Event Type  malfunction  
Manufacturer Narrative
B3: unknown.One device was received for evaluation.Visual inspection found the tamper seal to be broken.Functional testing was able to verify and duplicate the reported issue.It was determined that the out of specification battery compartment was the root cause and replaced.The service history review identified no indication that the complaint was related to a service of the device within the review period.
 
Event Description
It was reported that the device's battery compartment was burnt.The product fault occurred during testing, there was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD SOLIS VIP PUMP
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan ln n
minneapolis, MN 55442
2247062300
MDR Report Key18996843
MDR Text Key338915096
Report Number3012307300-2024-01758
Device Sequence Number1
Product Code FRN
UDI-Device Identifier15019517126587
UDI-Public15019517126587
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111275
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2120
Device Catalogue Number21-2120-0104-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2024
Date Manufacturer Received02/28/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/20/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-