GYRUS ACMI, INC. G400 GENERATOR, GYRUS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
Model Number 777000 |
Device Problem
Output below Specifications (3004)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/27/2023 |
Event Type
malfunction
|
Event Description
|
It was reported, the generator's output was too low.The issue occurred during maintenance of a diagnostic procedure.There were no reports of patient harm.
|
|
Manufacturer Narrative
|
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 16 years since the subject device was manufactured.The device was returned to olympus for inspection, and the customer's allegation was confirmed.Based on the results of the investigation, it is likely a faulty circuit board led to the output issues.The cause of the board becoming faulty and the auxiliary board being broken is unknown.Olympus will continue to monitor field performance for this device.
|
|