Catalog Number CDS0706-NTW
Device Problem Incomplete Coaptation (2507)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/06/2024
Event Type  Injury  
Event Description
It was reported that a mitraclip procedure was performed to treat degenerative mitral regurgitation (mr) with a grade of 4+ and a pre-existing big stitch located on the a2 leaflet.A mitraclip xtw was inserted and deployed on the mitral valve.However, after deployment, the clip detached from the posterior leaflet and remained attached to the anterior leaflet (single leaflet device attachment/slda).To stabilize the slda clip, an additional clip was implanted, reducing the mr from torrential to grade 2+.There was no clinically significant delay in the procedure.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot-specific quality issue.Based on available information, the cause of the reported slda and difficult imaging were unable to be determined.The reported unexpected medical intervention was a result of case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18997180
MDR Text Key338883231
Report Number2135147-2024-01386
Device Sequence Number1
Product Code NKM
UDI-Device Identifier05415067037435
UDI-Public(01)05415067037435(17)240315(10)30317R1027
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/15/2024
Device Catalogue NumberCDS0706-NTW
Device Lot Number30317R1027
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/06/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER
Patient Outcome(s) Required Intervention;
Patient Age68 YR
Patient SexMale
-
-