• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1104
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/19/2024
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the ventricular assist device (vad) was not returned for evaluation.No performance allegations were made against the device; therefore, the purpose of this investigation is solely to address the finding identified during analysis of the controller log files.Review of the motor start report revealed motor start events in which the motor start parameters were atypical.As a result, the finding was confirmed.Based on a historical review of similar events, the most likely root cause of atypical motor start parameters may be attributed, but not limited, to outer shroud contact that creates more friction at the housing to impeller interface during pump startup.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
Retrospective data file review conducted on patients no longer on ventricular assist device (vad) support (b)(6) 2024 revealed multiple motor start events with parameters outside of the typical range.The vad is not in a subgroup population for delay or failure to restart.The vad remained in use.No patient complications were reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP
Type of Device
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18999290
MDR Text Key338899561
Report Number3007042319-2024-01381
Device Sequence Number1
Product Code DSQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/31/2015
Device Model Number1104
Device Catalogue Number1104
Was Device Available for Evaluation? No
Date Manufacturer Received03/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-