• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC DAVINCI XI; VISION SIDE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC DAVINCI XI; VISION SIDE SYSTEM Back to Search Results
Model Number 381121-40
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2024
Event Type  malfunction  
Event Description
It was reported that during a da vinci-assisted esophageal diverticulum repair surgical procedure, the system reflected an erbe integrated electro surgical unit (iesu) error c-00 and c-34 errors.The troubleshooting steps were unsuccessful.The isi tse checked logs and confirmed errors pointing to the erbe iesu.The isi tse asked the customer to hard power cycle erbe iesu, but errors persisted.The customer confirmed there was no external high frequency source nearby and the erbe iesu was using a dedicated, well-grounded power outlet.The customer would use an external generator to complete the procedure.The procedure continued as planned with no reported injury.
 
Manufacturer Narrative
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to investigate the reported event further.Based on the field evaluation, this reported event was confirmed.The isi fse replaced the integrated electro surgical unit (iesu) to resolve the issue.The system was tested and verified as ready for use.Isi has received the iesu involved with this complaint to perform failure analysis; however, the evaluation is not yet complete.A follow-up mdr will be submitted once the product is evaluated and/ or if additional information is received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAVINCI XI
Type of Device
VISION SIDE SYSTEM
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key19000209
MDR Text Key339010892
Report Number2955842-2024-12732
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874110898
UDI-Public(01)00886874110898
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K131861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial
Report Date 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number381121-40
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/11/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/28/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-