AIZU OLYMPUS CO., LTD. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE Back to Search Results
Model Number LTF-S190-10
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was observed that during the device evaluation, the adhesive around the objective lens was peeled on the videoscope.There were no reports of patient involvement.
 
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the event, ¿glue at objective lens peeled off¿, was confirmed, and the device did not meet the standard specifications.The probable cause was stress due to repeated use, external factors, or handling of the device.The operation manual describes how to inspect for the suggested event in ¿chapter 3 preparation and inspection 3.3 inspection of the endoscope¿ which could have detected the phenomenon.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19000321
MDR Text Key338993756
Report Number9610595-2024-06602
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170434662
UDI-Public04953170434662
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberLTF-S190-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2024
Was the Report Sent to FDA? No
Date Manufacturer Received03/21/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/17/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-