The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned and a valid serial number has not been provided.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Clinical data was reviewed and confirmed that freestyle libre sensors continue to be safe, effective, and perform as intended in the field.Stability data for freestyle libre sensors was reviewed and showed no anomalies or non-conformances that could have led to the complaint.A tripped trend review was conducted for the reported complaint and freestyle libre sensors, no trends were identified that would indicate any product related issues.If the product is returned, a physical investigation will be performed and a follow-up report submitted.The device mfg date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
A low readings issue with the abbott diabetes care (adc) device was reported.A caregiver reported that an unspecified low scan was received on the customer's adc device compared to unspecified readings obtained on a healthcare professional (hcp) meter.As a result, the customer experienced experienced sweating and a loss of consciousness.On (b)(6)2024, the customer was provided unknown treatment by a non-healthcare professional.The customer was then seen at a hospital where an unspecified laboratory result was obtained however, no further treatment was reported.There was no report of death or permanent impairment associated with this event.
|