Model Number DVBC3D4 |
Device Problems
Electrical /Electronic Property Problem (1198); Battery Problem (2885); Reset Problem (3019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/08/2024 |
Event Type
Injury
|
Event Description
|
It was reported that the implantable cardioverter defibrillator (icd) triggered end of service (eos) and exhibited a full power on reset (por).After the por, the right ventricular (rv) lead had high/undefined rv coil impedance and there was concern for a lead fracture.However, the pacing impedance and sensing were normal, and no noise was reproduced with patient maneuvers.Device data was reviewed by qualified engineering personnel which indicated that 3 pors occurred, and the high rv coil impedance was likely a miss-measurement caused by a very low battery voltage.The icd was explanted and replaced.The rv lead tested normal with no impedance issues noted on the analyzer and remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|