It was reported that a patient underwent a cardiac ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and reddish material was observed inside the pebax in addition, the electrode #2 was observed lifted.During the procedure, the catheter was moving and disappear during ablation.The medical team changed the cable with no success.The catheter was changed and the issue resolved.The procedure continued.No patient consequences were reported.
|
E1 initial reporter phone: (b)(6).The product investigation was completed.Device evaluation details: the device was returned to biosense webster (bwi) for evaluation.A visual inspection and magnetic sensor functionality test of the returned device were performed following bwi procedures.Visual analysis was performed, and reddish material was observed inside the pebax in addition, the electrode #2 was observed lifted.The device was connected to the carto 3 system, and it was recognized and visualized correctly.Scanning electrode microscope (sem) analysis was performed on the pebax area and evidence of separation between the sleeve and the electrode was observed.A manufacturing record evaluation was performed for the finished device [31089008l] number, and no internal action related to the reported complaint condition were identified.The reddish material inside the pebax could be related to the visualization issue reported by the customer; therefore, the customer complaint was confirmed.The potential cause of the damage on the pebax and the electrode could be related to the usage of the device during procedure; however, this cannot be conclusively determined.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer's reference number: (b)(4).
|