SENSEONICS, INCORPORATED EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 101967-950
Device Problems Premature End-of-Life Indicator (1480); Calibration Problem (2890); Chemical Problem (2893); Device Ingredient or Reagent Problem (2910)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/29/2023
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted retrospectively as part of internal review.The sensor was tested in-house, and the review of investigation analysis revealed a loss of chemical performance, which confirmed the complaint.The system correctly disabled the sensor when it detected the performance failure, and the system's self-test functions were working normally.The root cause of the performance failure was due to oxidation of the chemical component of the sensor.As part of resolution, a return material authorization was issued for sensor replacement.No further resolution was necessary for this complaint.
 
Event Description
On jul 31, 2023, senseonics was made aware of an incident where the user received an early sensor replacement alert resulting in an early sensor removal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS, INCORPORATED
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer (Section G)
SENSEONICS, INCORPORATED
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20876-7005
MDR Report Key19001504
MDR Text Key338936307
Report Number3009862700-2024-00480
Device Sequence Number1
Product Code QHJ
UDI-Device Identifier81749102330
UDI-Public81749102330
Combination Product (y/n)Y
Reporter Country CodePL
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/12/2023
Device Model Number101967-950
Device Catalogue NumberFG-4500-50-302
Device Lot NumberWP09042
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2023
Date Manufacturer Received07/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/26/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-