Catalog Number 100-00004
Device Problems Use of Incorrect Control/Treatment Settings (1126); Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/11/2024
Event Type  malfunction  
Manufacturer Narrative
The internal complaint file #(b)(4) has been logged for this incident for traceability.The device involved in this incident was not returned to belmont for investigation.This issue is obvious to the user as the patient's temperature values are continuously displayed on the screen.Other methods of thermoregulation can be used to warm the patient faster.Criticool setting was altered so that the re-warming would take place at the required times and not on the adult setting.Patient re-warmed with delay.The neonate was monitored throughout with skin and rectal temperature probes.Patient remains ventilated which is not a result of the delayed re-warming.The medical team and parents were informed.Within the criticool unit, patient mode is displayed in the top left section titled "patient".The selected patient mode is highlighted with a white box.In this scenario, adult mode is selected.To select a different mode, press the desired patient mode, either neonatal or adult.The new patient mode is highlighted with a white box.The user manual also contains the following note: "the neonatal and adult modes have different settings during operation".There was no report of patient injury as a result of the alleged incident.Belmont has contacted the user to request additional information and clinilogger data about the alleged incident.Without having the unit or clinilogger data for investigation, it is not possible to determine whether the unit malfunctioned or not.Without results of the device investigation, no conclusions can be drawn.A follow-up report will be submitted once the investigation is complete and additional information becomes available.
 
Event Description
Belmont received mir reference number (b)(4) which detailed the following; a neonate was programmed for rewarming on the criticool after being actively cooled.The rewarming process was expected to be within 6 hours but noted to have taken excessively longer as the neonate attained temperature approximately 16hours afterwards.The device was reviewed and noted to be on the 'adult' setting rather than the 'infant' /neonatal setting, this accounted for the delay.The device was suspected to be preset to the 'adult' setting which is not expected for a device that was designated for neonates.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRITICOOL
Type of Device
THERMAL REGULATING SYSTEM
Manufacturer (Section D)
BELMONT MEDICAL TECHNOLOGIES
780 boston road
billerica MA 01821
Manufacturer (Section G)
BELMONT MEDICAL TECHNOLOGIES
780 boston road
billerica MA 01821
Manufacturer Contact
urja jani
780 boston road
billerica, MA 01821
MDR Report Key19001523
MDR Text Key339334653
Report Number1219702-2024-00015
Device Sequence Number1
Product Code DWJ
UDI-Device Identifier07290012127038
UDI-Public7290012127038
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K083662
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Remedial Action Inspection
Type of Report Initial
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number100-00004
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/01/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-