The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd drive pcb.In addition, our technician confirmed that the segment fluid damage, the control body fluid damage, the down pulley wire fluid damage, the up pulley wire fluid damage, the left pulley wire fluid damage, the right pulley wire fluid damage, the mechanical base plate fluid damage, the electrical pin connector fluid damage, the lg connector fluid damage, the insertion flexible tube buckled, the light guide cable crushed, the operation channel (primary) leak, and the remote control buttons cut; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|