Heparin initiated with icu medical plum 360 pump with pe lined tubing.Rn returned to room 10-15 minutes after initiation and blood was backflowing from central line through a small hole/crack in the tubing.Medication and tubing removed and new line initiated.Tubing sequestered.Lot #9405818 implicated in 4 possible tubing defects at local facility and lot has been removed.Product complaint #: (b)(4).Polyethylene lined tubing, secure lock 107 inch, product 14248-28; lot either #13827210 or #9405818.Reference report #mw5153339, #mw5153341, #mw5153342.
|