Model Number N/A |
Device Problem
Device Dislodged or Dislocated (2923)
|
Patient Problem
Joint Dislocation (2374)
|
Event Date 02/13/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2024 - 01033.D10: cat# 650-1055 lot# 3087401 cer bioloxd option hd 28mm.Cat# 650-1064 lot# 3136909 cer option type 1 tpr sleve -6.The customer has indicated that the product will not be returned to zimmer biomet for investigation as it was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that patient underwent an initial right total hip arthroplasty.Subsequently, the patient was revised approximately 1 month later due to dislocation.It was reported that no further information is available.
|
|
Event Description
|
No further information at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing for the bearing.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|