• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RESONATE HF CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATORS (CRT-D)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RESONATE HF CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATORS (CRT-D) Back to Search Results
Model Number G547
Device Problems High impedance (1291); High Capture Threshold (3266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/18/2024
Event Type  malfunction  
Event Description
It was reported that the patient with this cardiac resynchronization therapy defibrillator (crt-d) exhibited high out of range pacing impedance measurements, measuring at 2272 ohms on the left ventricular (lv) channel.The impedance measurements were noted during the device interrogation.Upon further review, the threshold values were high as well.There was no noise reported.This device currently remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE HF CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATORS (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19003958
MDR Text Key338908585
Report Number2124215-2024-19059
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589744
UDI-Public00802526589744
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberG547
Device Catalogue NumberG547
Device Lot Number318682
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexMale
-
-