• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI CAPSUREFIX NOVUS LEAD MRI SURESCAN; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI CAPSUREFIX NOVUS LEAD MRI SURESCAN; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 5076-52
Device Problems Decreased Sensitivity (2534); Impedance Problem (2950)
Patient Problem Dyspnea (1816)
Event Date 03/16/2024
Event Type  malfunction  
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported that the patient experienced shortness of breath.The right atrial (ra) lead exhibited unipolar impedance and chronically low p waves were noted.The lead remains in use.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX NOVUS LEAD MRI SURESCAN
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key19003999
MDR Text Key339015116
Report Number2649622-2024-08693
Device Sequence Number1
Product Code DTB
UDI-Device Identifier00643169633759
UDI-Public00643169633759
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P930039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number5076-52
Device Catalogue Number5076-52
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/18/2024
Date Device Manufactured06/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexMale
-
-