BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Catalog Number D134805
Device Problem Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2024
Event Type  malfunction  
Event Description
It was reported that a patient underwent a pulmonary vein isolation (pvi) ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter.The physician wanted to push the catheter in the left atrium.Before this step, he wanted to flush the catheter and he realized that there comes no water out of the tip.They tried everything but the catheter was blocked.The water from flushing did not come out of the holes of the catheter even if they saw that the tubing kit and the pump was flushing correctly.They used a new catheter and the new catheter was able to get flushed normally.The procedure continued.No patient consequences were reported.
 
Manufacturer Narrative
The bwi product analysis lab received the device for evaluation.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Manufacturer Narrative
Per internal review on 1-apr-2024, it was determined that this event is not mdr-reportable.Bwi received information on 19-mar-2024, that indicated that no ablation occurred with the complaint device.This device was not used on the patient at all and therefore presents no patient risk.No further reports will be sent regarding this event.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key19004847
MDR Text Key339214783
Report Number2029046-2024-01052
Device Sequence Number1
Product Code LPB
UDI-Device Identifier10846835010183
UDI-Public10846835010183
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD134805
Device Lot Number31210778L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2024
Is the Reporter a Health Professional? No
Date Manufacturer Received04/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/09/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SMARTABLATE PUMP.
-
-