Catalog Number CDS0706-XT
Device Problems Difficult or Delayed Positioning (1157); Incomplete Coaptation (2507)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/06/2024
Event Type  Injury  
Manufacturer Narrative
The clip remains in the patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4, an anterior flail, and tethered posterior leaflet.It was noted imaging was challenging during the procedure.One xtw clip was successfully deployed on the mitral valve.To further reduce mr, an xt clip was inserted.It was noted to have been difficlut to grasp the leaflets, however, after various attempts, both leaflets were grasped and the clip was deployed.However, after deployment, the clip detached from anterior leaflet and remained attached to the posterior leaflet (single leaflet device attachment/slda).To stabilize the slda, an additional clip was implanted, reducing mr to a grade of 3-4.There was no clinically significant delay in the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL NORTHPOINT REG 3020950818
1820 bastian court
westfield IN 46074
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key19004877
MDR Text Key338929328
Report Number2135147-2024-01388
Device Sequence Number1
Product Code NKM
UDI-Device Identifier05415067037404
UDI-Public(01)05415067037404(17)240930(10)31002A2003
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCDS0706-XT
Device Lot Number31002A2003
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/06/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
IMPLANTED MITRACLIP; STEERABLE GUIDE CATHETER
Patient Outcome(s) Required Intervention;
Patient Age66 YR
Patient SexMale
Patient Weight57 KG
-
-