Model Number 799-AEB1-00-01 |
Device Problem
No Display/Image (1183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation for evaluation.The customer's report was not replicated or confirmed.The device was put through extensive testing without duplicating the report.An internal inspection of the device found no discrepancies.A cable from the cpu to pim board and the uim board were replaced as a precaution.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
|
Event Description
|
Complainant alleged that during training by facility staff, the device's display blanked out.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|