Model Number G447
Device Problems Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Under-Sensing (1661)
Patient Problem Shock from Patient Lead(s) (3162)
Event Date 03/17/2024
Event Type  malfunction  
Event Description
It was reported that this pacemaker delivered inappropriate anti-tachycardia pacing (atp) and shock therapy for supraventricular tachycardia (svt) as the atrial signals were undersensed.One shock was delivered.Technical services (ts) suggested reprogramming options.There were no changes to programming.The device remains in use.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE X4 CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112-5798
6515826168
MDR Report Key19006027
MDR Text Key338944502
Report Number2124215-2024-19077
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589539
UDI-Public00802526589539
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/08/2024
Device Model NumberG447
Device Catalogue NumberG447
Device Lot Number279494
Was Device Available for Evaluation? No
Date Manufacturer Received03/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/21/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age52 YR
Patient SexMale
-
-