MAUDE Adverse Event Report: ZOLL CIRCULATION INC AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION INC AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2024
Event Type  malfunction  
Manufacturer Narrative
Zoll has not received the autopulse platform in the complaint for investigation.A supplemental report will be filed if and when the product is returned and investigation has been completed.
 
Event Description
During the patient call, the crew attempted to use the autopulse platform (sn unknown) on a patient but were unable to utilize the platform.The customer assumes an error message was displayed but is unsure.The crew did manual cpr before and after the attempt to use the platform.They did cpr for 25 min on site and another 25 min in the hospital.No impact or consequence to the patient was reported.After the code, the platform was examined, and the customer reported the plastic clip of the autopulse lifeband (lot unknown) was damaged inside of the gear shaft of the autopulse platform.Please see the following related mfr report: mfr #3010617000-2024-00262 for the autopulse lifeband.
 
Manufacturer Narrative
D4 (additional device information) was updated.The reported complaint that the autopulse platform sn (b)(6) could not be utilized, possibly due to an error message was confirmed during the archive review but not during functional testing.According to the archive, the platform displayed user advisory (ua) 02 (compression tracking error) around the event date.Based on the archive review, the device stopped compressions multiple times due to (ua) 02.The archive revealed that the take-up target and the max load sum exceeded 43 lbs, confirming that the size of the patient/object is too small and light in weight.The load sensors did not detect the expected increase in load during the compression.However, no device malfunction was observed during the functional testing and the autopulse platform performed as intended.The secondary complaint that the plastic clip of the autopulse lifeband (lot unknown) was damaged inside of the gear shaft of the autopulse platform was not confirmed during visual inspection or functional testing.During the service evaluation, no broken lifeband plastic clip was found inside the gear shaft.No issues were noted during testing and the platform performed as intended.Visual inspection of the returned platform was performed, and no physical damage was observed.A review of the archive data showed user advisory (ua) 02 (compression tracking error), related to the complaint.The autopulse platform passed the initial functional test without any fault or error.The reported complaint and (ua) 02 seen in the archive could not be reproduced.Load cell characterization was performed, and the results indicated that both load cells function within the specification.The platform was tested with the regular manikin using known good test batteries until discharged without any faults or errors.User advisory is normally a clearable error message and is designed into the autopulse platform to alert the operator that autopulse has detected one of several conditions.Per the battery hangtag - advisory codes description and action, user advisory 2 is an indication that the autopulse® has detected a change in lifeband tension.This advisory can happen when the patient or lifeband is out of position, or if the lifeband is opened during active operation.The recommended actions to take for this type of user advisory are: ensure that the lifeband is properly closed.Pull up completely on the lifeband, ensure that both the patient and the band are properly aligned, and press restart.Following service, the autopulse platform passed the run-in test without any fault or error.The autopulse platform passed the final testing without any fault or error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION INC
2000 ringwood ave
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION INC
2000 ringwood ave
san jose CA 95131
Manufacturer Contact
kimthoa sackrison
2000 ringwood ave
san jose, CA 95131
4084192922
MDR Report Key19006060
MDR Text Key338980319
Report Number3010617000-2024-00261
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000017
UDI-Public00849111000017
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-